
Installing the Disposables
4-4
20077/021 US
X
TRA
Operator’s Manual
Figure 4-2 Positioning the Blood Collection Reservoir on the Reservoir Holder
Note:
A Blood Collection Reservoir can be also mounted on an IV pole. However, it requires the use
of the dedicated additional holder.
4.
Tighten all the caps and luer locks on the reservoir lid. In case you are using a Blood Collection
Reservoir BOTTOM, close the clamp on the bottom outlet port.
5.
Set up of the three-way adapter (Y adapter) for use in cardiac surgery (if you are not using it, you
can skip next steps).
If the Blood Collection Reservoir is used during cardiac surgery or has to be connected to a
second reservoir, connect the Y adapter, provided in the Cardio Kit (refer to “Vacuum Line” on
page 3-20), to the reservoir outlet port . . .
a. located on the reservoir lid, marked WASH and covered with a blue cap, for a Blood Collection
Reservoir TOP
b. located on the bottom of the reservoir for a Blood Collection Reservoir BOTTOM.
6.
Close the clamps on the Y adapter branches waiting for subsequent connections. The branch
ending with a male port is used to connect the reservoir to the X
TRA
Bowl Set, the second one to
the Oxygenator Extension Line.
Step 2: Connecting the Suction Line
1.
Raise the IV pole completely.
2.
Prepare a bag or bottle of anticoagulant solution by mixing 30,000 international units of heparin
per liter of sterile (injectable) normal saline.
Note:
Premixed ACD-A or CPD solutions may be used in place of the heparin solution.
3.
Remove the dual lumen suction line from its package and carefully open the outer wrap.
WARNING
Before use, check that the Blood Collection Reservoir is completely
inserted in its housing on the holder. In the event of incorrect
installation, any accidental knocks against the reservoir might cause
it to detach from the holder.
CAUTION
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...