Preoperative Sequestration (PPP and PRP)
X
TRA
Operator’s Manual
20077/021 US
10-5
Setup for Preoperative Sequestration
The setup for preoperative sequestration differs slightly from that for intraoperative blood recovery.
When using the sequestration protocols, you will first install disposables and set up for sequestration.
When the sequestration procedure is completed, it is easy to convert the system quickly to
intraoperative blood recovery.
Supplies Required
The following supplies are required to set up the X
TRA
for both sequestration and intraoperative blood
recovery.
1.
X
TRA
Sequestration Set
2.
X
TRA
Bowl Set with 55 ml, 125 ml, 175 ml, or 225 ml bowl
3.
X
TRA
Collection Set
WARNING
WARNING
To reduce risk of air embolism, remove all air from the plasma bags
before handing the bag over for reinfusion.
WARNING
WARNING
i
American Association of Blood Banks.
Standards for Perioperative Autologous Blood
Collection and Administration
. 3rd Edition. Bethesda, MD. 2007 Reference Standard
5.1.8 (Handling, Storage, Transportation)
CAUTION
Make sure that you attach both PRP bags included in the X
does not alert you when a specified volume is
collected, you must have sufficient bags available for the entire amount that
CAUTION
Store plasma products containing platelets at room temperature (20 °C to 24
°C) with continuous, gentle agitation.
i
i
American Association of Blood Banks. Standards for Blood Banks and Transfusion
Services. 18th ed. Bethesda, MD: American Association of Blood Banks, 1997:
G3.300.
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...