Processing
5-48
20077/021 US
X
TRA
Operator’s Manual
3. Save Modifications button
The Save Modifications button appears to the left of the Close button as soon as any
modifications are made to the current protocol or operating mode. Touching the
Modifications Button applies all modifications and closes the Menu Screen.
4. Close button
Touching the Close button will close the “Protocol/Mode” tab without saving or applying any
modifications made to the operating protocol or mode.
Settings Tab
The Settings tab of the Menu Screen is used to view and modify the current configuration of the X
TRA
The settings of the current configuration may be viewed at any time (by touching the “View” button);
however, settings may only be modified if the Menu Screen was accessed from the Setup Screen (no
changes to the configuration settings may be made after the X
TRA
system has been loaded). The
majority of settings can be modified only by accessing the password protected area; a few changes
are not password protected to allow eventual modifications for the current case, before loading the
pump. All changes made from the Settings tab take effect immediately. The “Anticoagulant”,
“ReservoirType”, and “Vacuum” settings affect only the current case.
Figure 5-46 Settings Tab of the Menu Screen
“View” Button
Touching the “View” button allows the operator to view all of the settings in the current
configuration, but does not allow them to be modified. Use the “Configuration Mode”
button to modify the configuration settings.
Anticoagulant
The “Anticoagulant” setting may be used to toggle the anticoagulant type between “Heparin” and
“ACD”. The currently selected setting appears inset and green.
The set anticoagulant type is used by the software’s hematocrit signal conversion algorithms.
Changing the anticoagulant type here will affect the hematocrit values determined by the system
during processing.
Reservoir Type
The “ReservoirType” setting may be used to toggle the reservoir type between “X
TRA
” and “none”. The
currently selected setting appears inset and green.
When “ReservoirType” is set to “X
TRA
”, the system software will take into consideration the
characteristics of the standard X
TRA
reservoir in order to manage the functions related to it (such as
the Autostart function and the “Reservoir Full” alarm). This is the preferred setting when using the
standard X
TRA
reservoir.
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...