Troubleshooting
14-10
20077/021 US
X
TRA
Operator’s Manual
Bowl did not empty
completely
Warning
The bowl did not empty
completely.
• Press EMPTY to resume the
current phase. Reduce pump
speed, if necessary.
• Press X to exit and display the
ready screen.
WARNING: Repeating the bowl
emptying several times may
lead to elimination of all
residual air and possible
collapsing of tubings and waste
bag, and inflation of the RBC
bag.
If the problem persists, call the
authorized technical service.
Bowl not empty
Bowl did not empty
completely: do not
start the new cycle
before emptying the
bowl
Warning
The rapid transfer fill and empty
phases are based on volumes.
Resuming the fill phase with
volume still present in the bowl
may result in spillage of blood
into the waste bag.
• Reduce pump speed, if
necessary. Press EMPTY to
resume the Empty phase.
• Press X to exit and display the
ready screen.
WARNING: Repeating the bowl
emptying several times may
lead to elimination of all
residual air and possible
collapsing of tubings and waste
bag, and inflation of the RBC
bag.
Bowl not empty (RT)
Bowl not filled
Warning
The bowl is not completely
filled.
• Press PLAY to resume the
current phase.
• If the bowl is sufficiently
packed with RBCs, you may
decide to press WASH to
proceed to the Wash phase or
2xWASH to use twice the
saline solution volume.
WARNING: Inadequate filling
may lead to lower than
expected hematocrit and
removal of waste components.
Bowl not filled
Message Text
Warning or
Alarm
Buttons Available in the
Button Bar During
Warning/Alarm
Operator Response
Data Management
Text
Table 14-2 List of Warnings and Alarms Which Might Occur During Processing (Continued)
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...