Troubleshooting
14-4
20077/021 US
X
TRA
Operator’s Manual
Centrifuge test failed
Warning
The centrifuge could not
rotate properly.
Do not force the centrifuge
lid.
• Verify that the centrifuge
is not blocked by any
object, which might
prevent its proper
rotation. Press PLAY to try
to resume.
• Restarting the machine
might solve the problem.
If the problem persists, call
the authorized technical
service.
Centrifuge test failed
Check centrifuge arm
closure
Warning
The centrifuge arm might
be mispositioned.
Check bowl insertion into
the centrifuge, verify that
the centrifuge arm is
locked, close the centrifuge
lid and press PLAY to
resume.
If the problem persists, call
the authorized technical
service.
Check centr. arm closure
Check centrifuge lid lock
Warning
The centrifuge lid lock is
malfunctioning.
Press X to restore Lock
position, then close the lid.
If the problem persists, call
the authorized technical
service.
Check centr. lid lock
Current bowl size
different from previous
Warning
You are trying to retain data
but the bowl you have just
mounted is of a different
size from the previous one.
• If you want to start a new
case, press LOAD PUMP.
• If you want to continue
the last case with a
different bowl size, press
LOAD PUMP and reopen
the case from the Tally
screen.
Bowl size different
Message Text
Warning
or Alarm
Buttons Available in the
Button Bar During
Warning/Alarm
Operator Response
Data Management Text
Table 14-1 List of Warnings Which Might Occur During Setup (Continued)
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...