Preoperative Sequestration (PPP and PRP)
10-8
20077/021 US
X
TRA
Operator’s Manual
Connecting the Blood Bag Connection Line
11. Remove the blood bag connection line from the package. If necessary, connect the adapter with
spike
12. Connect the blood bag connection line to the inlet port of the bowl set
13. Spike or connect the whole blood bag with the blood bag connection line
14. Unclamp the lines as appropriate
Figure 10-5 Connecting the Blood Bag Connection Line
Selecting the Active Protocol and Operating Mode
Each cycle of a preoperative sequestration case may be processed using a different sequestration
protocol and operating mode (Automatic or Manual). The protocol may be selected only before the
cycle is initiated. The operating mode may be changed at any time during processing. The protocol
and mode are selected from the Protocol/Mode tab of the Menu Screen.
The process of selecting the active sequestration protocol and operating mode is the same as for
selecting an intraoperative or postoperative protocol and operating mode. For step-by-step
instructions refer to “Running a Case With the X
WARNING
Positioning the collection bags with the inlet line facing down does
not permit the correct execution of the sequestration procedure.
CAUTION
Use aseptic technique when installing disposables.
CAUTION
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...