Operator’s Manual
During Processing
Table 14-2 lists the warnings and alarms which may occur during processing with the X
TRA
. Each
warning or alarm is listed in alphabetical order by its message text.
If the warning or alarm persists after performing the actions listed in the “Operator Response” column,
call the authorized technical support service.
Message Text
Warning or
Alarm
Buttons Available in the
Button Bar During
Warning/Alarm
Operator Response
Data Management
Text
Abnormally long fill
cycle
Alarm
More than default fill volume
has been pumped into the
bowl.
• If the bowl is not sufficiently
packed with RBCs, press PLAY
resume the current phase.
• If the RBCs are spilling out of
the bowl, the RBC sensor
might be faulty and you might
want to proceed in MANUAL
MODE (If enabled).
If so, press STOP, enter the
MENU screen and, from the
PROTOCOL MODE tab, set
the protocol to Pstd and
switch to MANUAL MODE
• You may decide to disable the
alarm for the current case by
pressing DISABLE ALARM.
If the problem persists, call the
authorized technical service.
Fill cycle too long
Abnormally long fill
cycle (PPP/PRP)
Alarm
More than default fill volume
has been pumped into the
bowl. The PPP/PRP
sequestration could be
compromised.
• If the bowl is not sufficiently
packed with RBCs, press PLAY
resume the current phase.
• If the RBCs are spilling out of
the bowl, the RBC sensor
might be faulty and you might
want to proceed in MANUAL
MODE (If enabled).
If so, press STOP, enter the
MENU screen and, from the
PROTOCOL MODE tab, switch
to MANUAL MODE
• You may decide to disable the
alarm for the current case by
pressing DISABLE ALARM
If the problem persists, call the
authorized technical service.
Fill cycle too long, PPP/
PRP
Table 14-2 List of Warnings and Alarms Which Might Occur During Processing
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...