Vacuum Module
X
TRA
Operator’s Manual
20077/021 US
13-1
Chapter 13: Vacuum Module
Description
The X
VAC
is an optional vacuum module that can only be used as an accessory to the X
TRA
device. It is
available to provide negative vacuum pressure (-10 mmHg to -300 mmHg [-1.3 kPa to -40 kPa]). It
includes a reusable fluid receptacle and a replaceable hydrophobic filter which can be removed without
special tools.
Figure 13-1 X
VAC
Vacuum Module
The vacuum module has to be installed into the X
TRA
cart (see Figure 13-2), where it is powered by
the X
TRA
power supply and controlled using the X
TRA
touch screen user interface.
The X
VAC
vacuum module can be controlled using its built-in control panel only when the touch screen
user interface of X
TRA
is not available, in the following two cases:
1.
When the X
TRA
system is booting up
2.
When a fatal error occurred on the X
TRA
system
As soon as the touch screen user interface of X
TRA
is available, the X
VAC
vacuum module operates in
remote control, through the X
TRA
touch screen user interface.
The X
VAC
module has two operating modes:
• Intraoperative (Intra)
• Postoperative (Post-op)
The pressure set point for each mode may be set independently:
• Intraoperative mode: -300 mmHg to -50 mmHg (-40 kPa to -6.6 kPa)
• Postoperative mode: -100 mmHg to -10 mmHg (-13.3 kPa to -1.3 kPa)
In both modes, the pressure may be increased and decreased in -10 mmHg (-1.3 kPa) steps.
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...