Programmability Option
9-4
20077/021 US
X
TRA
Operator’s Manual
Figure 9-2 Fill Phase Screen With Modifiable Parameter
For a detailed explanation of the
Fill phase screen, read “Fill
Screen” on page 5-39.
The number and type of modifiable parameters depend on the selected protocol. Below is a table
summarizing the parameters of each factory protocol (or any custom protocol derived from it) that
may be modified at run time during the appropriate phase.
Changes to parameters made through the processing displet of the current phase take place
immediately: while the modification is displayed it is also put into effect.
Modifications performed during the current phase are not persistent across cases. When a new case is
started (from the Setup Screen), parameters will be restored to the values they had before the
modifications were applied. This reversion to default values applies to both factory and custom
protocols.
Modifications to the Popt, Pfat and Post-op protocol’s wash pump speed, wash volume, and emptying
pump speed will cause an asterisk to appear next to the protocol’s name as an indication to the
operator that the protocol has been modified.
Additionally, modifications to the Final Fill and Spill PRP pump speed parameter in the PRP protocols
have a life-span limited to only the current cycle. At the start of a new cycle, the factory values will be
set again.
Phase/Cycle
Popt
Pstd
Pfat
Post-op
PPP
PRP1
PRP2
Fill
Fill Pump Speed
Fill Pump Speed Fill Pump Speed Fill Pump Speed
Final Fill Pump
Speed
Final Fill Pump
Speed
Final Fill Pump
Speed
Wash
Wash Pump
Speed
Wash Pump
Speed
Wash Pump
Speed
Wash Pump
Speed
Wash Volume
Wash Volume
Wash Volume
Wash Volume
Spill
Spill PPP Pump
Speed
Spill PPP Pump
Speed
Spill PPP Vol.
Spill PPP Vol.
Spill PRP Pump
Speed
Spill PRP Pump
Speed
Spill PRP Vol.
Spill PRP Vol.
Empty
Empty Pump
Speed
Empty Pump
Speed
Empty Pump
Speed
Empty Pump
Speed
Empty Pump
Speed
Empty Pump
Speed
Empty Pump
Speed
Concentrate
Concentrate
Pump Speed
Concentrate
Pump Speed
Concentrate
Pump Speed
Concentrate
Pump Speed
Return
Return Pump
Speed
Return Pump
Speed
Return Pump
Speed
Table 9-1 Parameters of Factory Protocols Modifiable at Run Time
1
1 Fill pump speed
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...