Special Cycles
6-4
20077/021 US
X
TRA
Operator’s Manual
hematocrit of the incoming blood (%). The Concentrate pump speed (ml/min.) may be adjusted by
touching the pump rotor text box.
Touching the Pause button will stop the pump rotor. The centrifuge will continue to
spin for five minutes. While the system is paused, the operator may use the Phase
buttons in the button bar to abort the Concentrate cycle and continue processing in
the selected (or a different) phase. To resume the paused Concentrate cycle, touch the Play button.
When the buffy coat is detected, exit from the Concentrate cycle comes about in the same way as in
the Fill or Spill phases, according to the active operating mode and protocol (refer to “Running a Case
With the X
System” on page 5-5 and “Running a Preoperative Sequestration Case” on page 10-9).
Note:
If the Last Bowl function is activated and the air sensor detects the RBC bag to be empty
before the bowl is filled, the system will stop and prompt the operator to decide how to continue
processing (refer to “Last Bowl” on page 7-4).
Priming the Reinfusion Line: Prime IV
The X
TRA
system has a Prime IV function which can be used to pump 5 ml and 30 ml of wash solution
into the RBC bag and empty line, respectively, to simplify priming the reinfusion line. However, it is
strongly recommended that this feature not be used unless the reinfusion line is routed through
reinfusion protection devices. See “Reinfusion” on page 6-7 for instructions. The prime volume is not
counted in any tally data for the case.
The Prime IV cycle is initiated by touching the “Prime IV” button, which is accessible from
the Special Cycles Screen when the machine is stopped prior to the start of a processing
protocol (refer to “The Special Cycles Screen” on page 6-1). The Prime IV button is only
available if the function is enabled in the “Features” tab of the Configuration Mode Screen.
During this cycle, the Wash clamp is opened (the other clamps remaining closed) and the pump
rotates counterclockwise to move 47 ml saline solution from the Wash bag into the bowl. Then, the
Wash clamp is closed and the Empty clamp is opened, and the pump rotates clockwise to move 35 ml
CAUTION
The Concentration function reprocesses already collected red blood cells
subjecting them again to the mechanical action of the pump and the
centrifuge. The repeated use of the Concentration function on the same red
blood cells might lead to them being damaged and therefore to their loss.
WARNING
WARNING
Do not reinfuse the patient’s blood from the primary RBC bag when it is
connected to the X
autologous transfusion circuit. Reinfusion from
the primary reinfusion bag connected to the circuit could lead to air
embolism.
WARNING
WARNING
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...