System Description
3-12
20077/021 US
X
TRA
Operator’s Manual
Processing Pump
The processing pump (see Figure 3-16) moves fluids into and out of the centrifuge bowl while
providing high flow rates with minimal hemolysis. In the autotransfusion protocols, the processing
pump can be operated at rates of 25 to 1,000 ml/min. in 25 ml/min. increments. In both the PPP and
PRP protocols, the pump can be operated at rates of 10 to 100 ml/min. in 10 ml/min. increments.
Figure 3-16 Visible Parts of the X
TRA
Processing Pump
The processing pump can be unfixed from its seat for cleaning purposes, as described in
Air Detector
The air detector is located on the top console between the processing pump and the centrifuge bowl
(see Figure 3-17).
Figure 3-17 Location of the Air Detector
The air detector is an ultrasonic device that detects air in the tubing from the pump to the centrifuge
bowl. It is responsible for detecting the end point of the Empty phase and detecting when the
collection reservoir or wash bag becomes empty.
2
1 Pump rotor
2 Pump handle
1
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...