Troubleshooting
14-22
20077/021 US
X
TRA
Operator’s Manual
Weight not reliable.
Autostart
temporarily
disabled.
Warning
The weight of the reservoir is
compromised and the Autostart
automation has been
temporarily disabled.
No operator response.
Weight not reliable
Whole blood bag
empty.
Warning
The sensor is detecting air in
the fill line.
• The fill line might be clamped
or occluded. Remove the
occlusion and press PLAY to
resume.
• The whole blood bag might be
empty. The spill phase cannot
be started if the bowl has not
been properly filled with RBCs
(in PRP1 and PRP2). If you
intend to continue the
sequestration, connect a new
whole blood bag and press
PLAY to resume.
• If you intend to end the
sequestration, press EMPTY
to recover the bowl content
into the RBC bag.
If the problem persists, call the
authorized technical service.
Whole blood bag empty
Message Text
Warning or
Alarm
Buttons Available in the
Button Bar During
Warning/Alarm
Operator Response
Data Management
Text
Table 14-2 List of Warnings and Alarms Which Might Occur During Processing (Continued)
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...