Processing
X
TRA
Operator’s Manual
20077/021 US
5-31
The Operating Modes
The X
TRA
system is equipped with three operating modes:
• Automatic
• 1 Touch
• Manual
Each is described in the sections below.
Modes of operation influence the execution of the active protocol, affecting the transition between
phases and cycles.
The current operating mode can be set from the Protocol/Mode tab of the Menu Screen. The operating
mode may be changed while the machine is stopped (at the Ready Screen), or any time during a
cycle. Step-by-step instructions can be found in the section “Running a Case With the X
Automatic
In Automatic mode the system recognizes the end of each phase and automatically switches to the
next phase without any intervention from the operator (Fill to Wash/Spill to Empty). By default, at the
end of each cycle the system displays the Ready Screen and does not begin a new cycle automatically.
Automatic mode is compatible with all protocols and can be switched to at any time.
While Automatic mode is active, the word “Automatic” appears in the status area of every screen.
If an intraoperative or postoperative protocol is selected (Pstd, Popt, Post-op and Pfat), then either or
both of the Continue and Autostart functions may be enabled in conjunction with Automatic mode. If
both Continue and Autostart are enabled, then Automatic mode behaves exactly like 1 Touch mode.
1 Touch
The 1 Touch operating mode is a shortcut to running the Automatic mode with both the Autostart and
Continue functions enabled. It is only available if processing with an intraoperative or postoperative
protocol (Pstd, Popt, Pfat or Post-op).
In 1 Touch mode, each cycle is started automatically (Autostart function), the system recognizes the
end of each phase and automatically switches to the next phase without operator intervention (Fill to
Wash to Empty), and the system starts a subsequent cycle automatically as long as sufficient blood
volume is available in the reservoir (Continue function).
While 1 Touch mode is active, “1 Touch” appears in the status area of every screen.
The 1 Touch mode will not be available if disabled in the current configuration (refer to
).
Although it is possible to disable the Autostart function when 1 Touch mode is selected, it is considered
exceptional. On the contrary, it is not possible to disable the Continue function when 1 Touch mode is
selected.
Wash Volume
0 - 5000
100
ml
200
800
1000
1000
Empty Flow
25 - 1000
25
ml/min
150
300
400
400
Concentrate flow
25 - 1000
25
ml/min
200
350
450
300
i
These parameters can not be modified by the user.
ii
These parameters can be modified by the user but in case of modifications a marker will appear telling factory protocol has been modified.
Protocol
Parameters
Range
Step
Unit
Factory values
X/55
X/125
X/175
X/225
Table 5-2 ATS Protocol Parameter Summary (Continued)
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...