Preoperative Sequestration (PPP and PRP)
10-26
20077/021 US
X
TRA
Operator’s Manual
Figure 10-15 Spill PRP Phase of the Processing Screen
1. “Conc” Button
Touching the “Conc” button initiates the Concentrate cycle (refer to “The Concentrate Cycle
(With the PPP/PRP Protocols)” on page 10-16). This button is only present while processing
is paused.
2. Pump Speed
The pump speed (ml/min.) is displayed in a text box in the processing displet to the right of the
pump rotor icon. The pump speed may be adjusted by touching the text box and using the up
and down arrow buttons (refer to “Modification of Parameters” on page 5-5).
3. Spill Status Line
The spill status line consists of three elements, from left to right: an icon indicating the current
sub-phase (yellow for PPP Spill, orange for PRP Spill), a text field displaying the current volume
of PPP or PRP spilled, and a text box displaying the target volume of PPP or PRP to spill. The
target spill volume may be modified by touching the text box and using the up and down arrow
buttons.
“Spill PRP” Text Box
The “Spill PRP” field and text box displays the target volume of PRP to spill during the PRP Spill
sub-phase. The target spill volume may be modified by touching the text box and using the up and
down arrow buttons. The “Spill PRP” text box is only present during the PPP Spill sub-phase.
“Total Spill” Field
The “Total Spill” field displays the total volume (ml) of both PPP and PRP collected. It is only present
during the PRP Spill sub-phase.
4. Play/Pause button
Touching the Pause button pauses the processing cycle: the pump is stopped, the
Pause button changes to the Play button, and the word “Paused” flashes and a 5:00
countdown timer appears in the title bar of the processing displet. The centrifuge will
continue to spin for five minutes.
If Automatic mode is active, phase buttons appear in the button bar of the processing screen while the
cycle is paused. This allows for manually intervention in the phase progression. (If Manual mode is
active, these phase buttons are always visible.)
Touching the Play button resumes processing from the state the phase was in when it was paused.
Pausing and resuming a cycle is discussed further in “Pausing and Resuming the Cycle” on page 5-11.
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...