Disconnect the
patient from the
machine
Alarm
You are about to switch from a
postoperative protocol to an
intraoperative protocol. The
patient must be disconnected
from the machine.
Disconnect the patient before
proceeding. Once done, touch
the V button.
Warning
: Not following these
instructions may expose the
patient to high suction levels.
Disconnect the patient
Empty line occluded
Alarm
The empty line might be
clamped or occluded.
• Remove the clamp on the line
and/or verify the absence of
occlusions, then press PLAY
to resume the current phase.
• You may decide to disable the
sensor and the related alarm
for the current case by
pressing DISABLE ALARM.
If the problem persists, call the
authorized technical service.
Empty line occluded
Fluid loss in
centrifuge – Alarm
disabling
Warning
You are about to disable the
Centrifuge Fluid Loss Sensor
and related alarm.
Press DISABLE ALARM to
disable the alarm for the rest of
the case, or press X to maintain
it.
WARNING: Disabling this alarm
may result in loss of blood or
biohazard risk to operator. The
device must be directly
monitored by the operator until
the end of the case.
Message Text
Warning or
Alarm
Buttons Available in the
Button Bar During
Warning/Alarm
Operator Response
Data Management
Text
Table 14-2 List of Warnings and Alarms Which Might Occur During Processing (Continued)
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...