Processing
X
TRA
Operator’s Manual
20077/021 US
5-39
5. Emergency Button
Touching the Emergency button, and then confirming in the subsequent confirmation screen,
initiates the Emergency function (refer to “Emergency” on page 5-26).
While the Emergency function is active, this button appears in its pressed state (inset and green).
Touching it again will disable the Emergency function, re-enabling the previous protocol (confirmation
not required).
The Emergency button is only available during intraoperative and postoperative protocols.
6. “End Case” Button
Touching the “End Case” button displays the End of Case Screen and allows the operator to
end the current case (refer to “End of Case Screen” on page 5-50).
Processing Screen
Once a processing cycle has been initiated from the Ready Screen, the Processing Screen is displayed
on the touch screen display.
From the Processing Screen, each phase of each cycle may be monitored and controlled. The
processing displet (in the center of the screen) changes during processing to present controls and
information relevant to the current phase. The title bar of the processing displet always indicates the
current phase. In this section, the Processing Screen is described as four different screens, one for
each phase: Fill, Wash, Empty and Fat Removal (only present if Pfat protocol is being used).
Fill Screen
Figure 5-37 Fill Phase of the Processing Screen
1. Pump Speed
The pump speed (ml/min.) is displayed in a text box in the processing displet to the right of the
pump rotor icon. Except during the Fill phase of the Popt, Pfat and Post-op protocols, the pump
speed may be adjusted by touching the text box and using the up and down arrow buttons
(refer to “Modification of Parameters” on page 5-5).
2. Fill Status Line
The fill status line indicates the current phase with the fill icon. To the right of the icon is a text
field which displays the amount of fluid which has been pumped into the bowl (ml) during the
current cycle.
1 Pump speed
2 Fill Status line
3 Play/Pause button
1
2
3
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...