Remote Control Commands
R&S
®
SMBV100A
674
Operating Manual 1407.6062.32 ─ 08
Example:
PM:STAT ON
activates PM modulation.
Usage:
SCPI confirmed
7.14.17 SOURce:POWer Subsystem
This subsystem contains the commands for setting the output level, level control and level
correction of the RF signal.
Other units can also be used instead of dBm:
●
by entering the unit directly after the numerical value (example
:POW 0.5V
)
●
by changing the DEFault unit in the
UNIT
system (see the command
.............................................................................674
..............................................................................675
.............................................................................675
.........................................................676
..................................................676
........................................................677
.............................................................677
.......................................................................678
.....................................................................................678
.....................................................................................679
.......................................................................................680
..........................................................................................680
......................................................................................680
............................................................................681
...............................................................................681
................................................................................682
..............................................................................682
...............................................................................682
.............................................................................682
.......................................................................................683
....................................................................683
.............................................................................683
........................................................................................684
[:SOURce<hw>]:POWer:ALC:OMODe
<OffMode>
The command sets the level control mode which becomes active when automatic level
control is deactivated (ALC Off).