![Rohde & Schwarz SMBV100A Скачать руководство пользователя страница 575](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_manual_1477734575.webp)
Remote Control Commands
R&S
®
SMBV100A
558
Operating Manual 1407.6062.32 ─ 08
Example:
BB:DM:TRIG:SOUR EXT
sets external triggering via the TRIGGER 1 connector.
BB:DM:TRIG:MODE ARET
selects the Armed_Retrigger mode
BB:DM:TRIG:RMOD?
queries the current status of signal generation.
Response: RUN
the signal is generated, an external trigger was executed.
Usage:
Query only
[:SOURce<hw>]:BB:DM:TRIGger:SLENgth
<SLength>
The command defines the length of the signal sequence to be output in the "Single" trigger
mode.
Parameters:
<SLength>
float
Range:
1 to 4294967295
*RST:
1000
Default unit: symbol
Example:
BB:DM:SEQ SING
sets trigger mode Single.
BB:DM:TRIG:SLEN 200
sets a sequence length of 200 symbols. 200 symbols will be output
after the next trigger event.
[:SOURce<hw>]:BB:DM:TRIGger:SOURce
<Source>
The command selects the trigger source.
Parameters:
<Source>
INTernal | EXTernal
INTernal
Triggering is executed by means of the Trigger command
SOURce:BB:DM:TRIGger:EXECute
,
TRIGger:BB:DM:IMMediate
or
*TRG
in the case of remote
control and by means of "Execute Trigger" in the case of manual
operation.
EXTernal
Triggering is executed by means of the signal on the TRIGGER 1
connector.
Example:
BB:DM:TRIG:SOUR EXT
sets external triggering via the TRIGGER 1 connector.
SOURce Subsystem
Содержание SMBV100A
Страница 31: ......
Страница 117: ...Manual Operation R S SMBV100A 100 Operating Manual 1407 6062 32 08 Legend of Front Panel Controls ...