Instrument Function
R&S
®
SMBV100A
175
Operating Manual 1407.6062.32 ─ 08
The power-on behavior of the instrument and the level display in the display header are
set in the "Power-On / EMF Settings" section (see
chapter 5.4.8.3, "Power-On/EMF Set-
The lower sections provide access to the "Automatic Level Control" settings and to func-
tion User Correction settings. For more information about this functions, see
ter 5.4.9, "Automatic Level Control - ALC"
Level Settings
The offset-free level, attenuation mode, level offset and level limit are set in the top section
of the menu.
RF Mode
Selects performance preferences of the RF output signal regarding wideband noise and
distortion.
"Normal"
The generator provides an RF output signal with high signal to noise
ratio as well as low distortion, according to the data sheet.
"LOW Noise"
This setting forces the generator to optimize the signal to noise ratio.
"LOW Distor-
tion"
The generator reduces distortion (harmonics) of the RF signal.
SCPI command:
Amplitude
Sets the RF level of the RF output connector.
The level entered and displayed here corresponds to the level at the RF output, i.e. any
offset entry is not taken into consideration.
Note:
The SCPI command
sets the level of the "Level" display, i.e. the level containing offset.
SCPI command:
RF Signal and Analog Modulations - RF Block
Содержание SMBV100A