Remote Control Commands
R&S
®
SMBV100A
534
Operating Manual 1407.6062.32 ─ 08
[:SOURce<hw>]:BB:ARB:MCARrier:SETTing:STORe:FAST
<Fast>
Determines whether the instrument performs an absolute or a differential storing of the
settings.
Enable this function to accelerate the saving process by saving only the settings with
values different to the default ones.
Note:
This function is not affected by the "Preset" function.
Parameters:
<Fast>
0 | 1 | OFF | ON
*RST:
1
[:SOURce<hw>]:BB:ARB:MCARrier:TIME
<Time>
The command sets the user-defined signal period. This setting is only possible for Signal
Period Mode User (
BB:ARB:MCAR:TIME:MODE USER
).
Parameters:
<Time>
float
Range:
0 to 1E9
Increment: 1E-9
*RST:
0
Default unit: s
Example:
BB:ARB:MCAR:TIME:MODE USER
selects Signal Period Mode User.
BB:ARB:MCAR:TIME 10 s
sets a signal period of 10 seconds
[:SOURce<hw>]:BB:ARB:MCARrier:TIME:MODE
<Mode>
The command selects the mode for calculating the resulting signal period of the multi
carrier waveform. The resulting period is always calculated for all carriers in the carrier
table irrespective of their state (ON/OFF).
Parameters:
<Mode>
USER | LONG | SHORt
LONG
The resulting signal period is defined by the longest I/Q file in the
carrier table. Shorter I/Q files are periodically repeated.
SHORt
The resulting signal period is defined by the shortest I/Q file in the
carrier table. Only the first part of longer I/Q files is used.
USER
The signal period can be set with command
SOUR:BB:ARB:MCARr:TIME
. Shorter I/Q files are repeated peri-
odically, and only the first part of longer I/Q files is used.
*RST:
LONG
SOURce Subsystem