Remote Control Commands
R&S
®
SMBV100A
605
Operating Manual 1407.6062.32 ─ 08
Return values:
<Maximum>
float
Range:
0 to 16777215
Increment: 1E-3
Example:
BB:MCCW:TRIG:OUTP:DEL:FIX ON
restricts the marker signal delay setting range to the dynamic
range.
BB:MCCW:TRIG:OUTP:DEL:MAX
queries the maximum of the dynamic range.
Response: 2000
the maximum for the marker delay setting is 2000 samples.
Usage:
Query only
[:SOURce<hw>]:BB:MCCW:TRIGger:OUTPut<ch>:DELay:MINimum?
The command queries the minimum marker delay for set-
ting
:BB:MCCW:TRIGger:OUTPut: DELay:FIXed ON
.
Return values:
<Minimum>
float
Range:
0 to 16777215
Increment: 1E-3
*RST:
0
Example:
BB:MCCW:TRIG:OUTP:DEL:FIX ON
restricts the marker signal delay setting range to the dynamic
range.
BB:MCCW:TRIG:OUTP:DEL:MIN
queries the minimum of the dynamic range.
Response:0
the minimum for the marker delay setting is 0 samples.
Usage:
Query only
[:SOURce<hw>]:BB:MCCW:TRIGger:OUTPut<ch>:MODE
<Mode>
The command defines the signal for the selected marker output.
SOURce Subsystem