Remote Control Commands
R&S
®
SMBV100A
510
Operating Manual 1407.6062.32 ─ 08
Parameters:
<Output>
0 | 1 | OFF | ON
ON
The signal calculation starts simultaneously with the external trig-
ger event but because of the instrument’s processing time the first
samples are cut off and no signal is outputted. After elapsing of
the internal processing time, the output signal is synchronous to
the trigger event.
OFF
The signal output begins after elapsing of the processing time and
starts with sample 0, i.e. the complete signal is outputted.
This mode is recommended for triggering of short signal sequen-
ces with signal duration comparable with the processing time of
the instrument.
*RST:
ON
Example:
BB:ARB:TRIG:SOUR EXT
sets external triggering.
BB:ARB:TRIG:EXT:SYNC:OUTP ON
enables synchronous output to external trigger
[:SOURce<hw>]:BB:ARB:TRIGger:OUTPut:DELay:FIXed
<Fixed>
The command restricts the marker delay setting range to the dynamic range. In this range
the delay can be set without restarting the marker and signal. If a delay is entered in
setting ON but is outside this range, the maximum possible delay is set and an error
message is output.
The setting always affects every marker.
Parameters:
<Fixed>
0 | 1 | OFF | ON
*RST:
0
Example:
BB:ARB:TRIG:OUTP:DEL:FIX ON
restricts the marker signal delay setting range to the dynamic
range.
[:SOURce<hw>]:BB:ARB:TRIGger:OUTPut<ch>:DELay
<Delay>
The command defines the delay between the signal on the marker outputs and the start
of the signals, expressed in terms of samples.
Command
:BB:ARB:TRIGger:OUTPut:DELay:FIXed ON
can be used to restrict the
range of values to the dynamic range, i.e. the range within which a delay of the marker
signals can be set without restarting the marker and signal.
SOURce Subsystem
Содержание SMBV100A
Страница 31: ......
Страница 117: ...Manual Operation R S SMBV100A 100 Operating Manual 1407 6062 32 08 Legend of Front Panel Controls ...