Remote Control Commands
R&S
®
SMBV100A
600
Operating Manual 1407.6062.32 ─ 08
[:SOURce<hw>]:BB:MCCW:TRIGger:EXTernal:SYNChronize:OUTPut
<Output>
(enabled for Trigger Source External)
Enables/disables output of the signal synchronous to the external trigger event.
See also section
"Sync. Output to External Trigger"
on page 380 for a detailed description
of the applications of this setting.
Parameters:
<Output>
0 | 1 | OFF | ON
ON
The signal calculation starts simultaneously with the external trig-
ger event but because of the instrument's processing time the first
samples are cut off and no signal is outputted. After elapsing of
the internal processing time, the output signal is synchronous to
the trigger event.
OFF
The signal output begins after elapsing of the processing time and
starts with sample 0, i.e. the complete signal is outputted.
This mode is recommended for triggering of short signal sequen-
ces with signal duration comparable with the processing time of
the instrument.
*RST:
1
Example:
BB:MCCW:TRIG:SOUR EXT
sets external triggering.
BB:MCCW:TRIG:EXT:SYNC:OUTP ON
enables synchrounous output to external trigger.
[:SOURce<hw>]:BB:MCCW:TRIGger:OBASeband:DELay
<Delay>
The command specifies the trigger delay (expressed as a number of samples)for external
triggering.
Parameters:
<Delay>
float
Range:
0 to 65535
Increment: 0.01
*RST:
0
Example:
BB:MCCW:TRIG:SOUR EXT
selects an external trigger via the TRIGGER 1 connector.
BB:MCCW:TRIG:DEL 200
sets a delay of 200 samples for the trigger.
[:SOURce<hw>]:BB:MCCW:TRIGger:OBASeband:INHibit
<Inhibit>
The command specifies the number of samples by which a restart is to be inhibited fol-
lowing a trigger event. This command applies only in the case of external triggering.
SOURce Subsystem
Содержание SMBV100A
Страница 31: ......
Страница 117: ...Manual Operation R S SMBV100A 100 Operating Manual 1407 6062 32 08 Legend of Front Panel Controls ...