![Rohde & Schwarz SMBV100A Скачать руководство пользователя страница 588](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_manual_1477734588.webp)
Remote Control Commands
R&S
®
SMBV100A
571
Operating Manual 1407.6062.32 ─ 08
Parameters:
<Shape>
LINear | COSine
*RST:
COSine
Example:
BB:DM:PRAM:SHAP COS
selects a cosine-shaped rise and fall for the transmitted power
edge.
[:SOURce<hw>]:BB:DM:PRAMp:SOURce
<Source>
The command sets the source for the power ramp control signals.
Parameters:
<Source>
INTernal
INTernal
The internal control signals LEV_ATT and BURST are used. Con-
trol signals are stored in dedicated lists. When remote control is in
use, the list to be used is defined, selected and enabled with the
aid of
SOURce:BB:DM:CLISt:...
commands.
*RST:
INTernal
Example:
BB:DM:PRAM:SOUR INT
enables the use of internal control signals for power ramp control.
[:SOURce<hw>]:BB:DM:PRAMp:TIME
<Time>
The command sets the power ramping rise time and fall time for a burst.
Parameters:
<Time>
float
Range:
0.25 to 16
Increment: 0.01
*RST:
1
Default unit: symbol
Example:
BB:DM:PRAM:TIME 2
sets a time of 2 symbols for the edges to rise and fall.
[:SOURce<hw>]:BB:DM:PRAMp[:STATe]
<State>
The command enables or disables power ramping.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
BB:DM:PRAM:STAT ON
switches power ramping on.