Remote Control Commands
R&S
®
SMBV100A
613
Operating Manual 1407.6062.32 ─ 08
[:SOURce]:BBIN:OLOad:HOLD:STATe?
This command queries an overload since the last reset for evaluating the measurement.
The Overload Hold state can be reset via the "Reset" command, or is reset automatically
by starting a new measurement ("Auto Level Set") or by setting new values (Crest Factor,
PEP, Level).
Return values:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
BBIN:MODE DIG
selects the digital baseband input mode.
BBIN:OLO:HOLD:STAT?
queries the overload state.
Response: 0
the AC/DC converter has not been overloaded.
Usage:
Query only
[:SOURce]:BBIN:OLOad:STATe?
This command queries the current overload state of the A/D converter.
Return values:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
BBIN:MODE DIG
selects the digital baseband input mode.
BBIN:OLO:STAT?
queries the overload state.
Response: 0
The AC/DC converter is not overloaded.
Usage:
Query only
[:SOURce]:BBIN:POWer:PEAK
<Peak>
This command enters the peak level of the external baseband signal relative to full scale
of 0.5 V (in terms of dB full scale).
Parameters:
<Peak>
float
Range:
-60 to 3.02
Increment: 0.01
*RST:
0
Default unit: dBfs
Example:
BBIN:POW:PEAK -2
enters the peak level of -2 dBfs.
SOURce Subsystem