Remote Control Commands
R&S
®
SMBV100A
540
Operating Manual 1407.6062.32 ─ 08
Example:
MMEM:CDIR '/var/smbv/user/waveform'
sets the default directory to /var/smbv/user/waveform.
BB:ARB:WSEG:CONF:SEL 'multi_sin'
creates the configuration file
multi_sin.inf_mswv
in the
default directory.
BB:ARB:WSEG:CONF:SEGM:APP 'sinus1'
includes waveform
sinus1.wv
as segment 1 in the configuration
file. The waveform must be available in the default directory.
BB:ARB:WSEG:CONF:SEGM:APP 'sinus2'
includes waveform
sinus2.wv
as segment 2 in the configuration
file. The waveform must be available in the default directory.
BB:ARB:WSEG:CONF:OFIL 'mseg1_2'
defines the file name
mseg1_2.wv
for the multi segment wave-
form.
BB:ARB:WSEG:CLO 'multi_sin'
creates multi segment waveform
mseg1_2.wv
using the settings
of the configuration file
multi_sin.inf_mswv
.
The new multi segment waveform is loaded and digital standard
"ARB" is activated.
Usage:
Setting only
[:SOURce<hw>]:BB:ARB:WSEGment:CONFigure:BLANk:APPend
<SampCount>,
<Frequency>
Adds a blank segment to the multi segment file.
Setting parameters:
<SampCount>
float
Specifies the number of samples.
Range:
512 to 1E7
Increment: 1
*RST:
1000
<Frequency>
float
Determines the clock rate.
Range:
400 Hz to Max (depends on instrument type)
Increment: 0.001
*RST:
1E8
Example:
SOUR:BB:ARB:WSEG:CONF:SEL "MSegFile"
selects a multi segment file.
SOUR:BB:ARB:WSEG:CONF:BLAN:APP 1000,100000000
adds a blank segment with 1000 samples and 100 MHz clock rate
to the selected multi segment file
Usage:
Setting only
SOURce Subsystem