Remote Control Basics
R&S
®
SMBV100A
406
Operating Manual 1407.6062.32 ─ 08
Instrument responses are displayed on the "viRead" tab.
Tip:
For further program operation refer to the online help of the program.
6.2.3.3
Remote Control over LAN using Socket Communication
This chapter provides an example on how to establish a remote control connection over
telnet protocol and a simple sockets-based program example that can be further devel-
oped.
Setting up a Telnet Connection
To control the software, only a telnet program is required. The telnet program is part of
every operating system.
1. To establish a Telnet connection with the R&S SMBV, start the telnet program and
enter the socket address.
The socket address is a combination of the IP address or the host name of the R&S
SMBV and the number of the port configured for remote-control via telnet.
Tip:
The R&S SMBV uses the port number 5025 for remote connection via Telnet.
The connection to the instrument is set up and remote-control commands can be
sent.
Starting a Remote Control Session