Instrument Function
R&S
®
SMBV100A
316
Operating Manual 1407.6062.32 ─ 08
Trigger In
The "Trigger In" section is used to configure the trigger signal for the digital modulation.
The current status of signal generation is indicated for all trigger modes.
Trigger Mode - Digital Modulation
Selects trigger mode.
"Auto"
The digital modulation signal is generated continuously.
"Retrigger"
The digital modulation signal is generated continuously. A trigger event
(internal with "Execute
Trigger" or external) causes a restart.
"Armed_Auto"
The digital modulation signal is generated only when a trigger event
occurs. Then the digital modulation signal is generated continuously.
Button "Arm" stops signal generation. A subsequent trigger event (inter-
nal with "Execute
Trigger" or external) causes a restart.
"Armed_Ret-
rigger"
The digital modulation signal is generated only when a trigger event
occurs. Then the signal is generated continuously. Every subsequent
trigger event (internal with "Execute
Trigger" or external) causes a
restart.
The button "Arm" stops signal generation. A subsequent trigger event
(internal with "Execute
Trigger" or external) causes a restart.
"Single"
The digital modulation signal is generated only when a trigger event
occurs. Then the signal is generated once to the length specified at
"Signal Duration". Every subsequent trigger event (internal with "Exe-
cute
Trigger" or external) causes a restart.
SCPI command:
Trigger Signal Duration - Digital Modulation
Enters the length of the signal sequence to be output in the "Single" trigger mode. The
input is to be expressed in symbols.
SCPI command:
Running - Stopped - Digital Modulation
Displays the status of signal generation for all trigger modes. This display appears only
when Custom Dig Mod is enabled (State On).
"Running"
The digital modulation signal is generated; a trigger was (internally or
externally) initiated in triggered mode.
If Armed_Auto and Armed_Retrigger have been selected, generation
of signals can be stopped with the "Arm" button. A new trigger (internally
with "Execute
Trigger" or externally) causes a restart.
"Stopped"
The signal is not generated, and the instrument waits for a trigger event
(internal or external).
SCPI command:
Baseband Signal - Baseband Block