Remote Control Commands
R&S
®
SMBV100A
690
Operating Manual 1407.6062.32 ─ 08
Options:
R&S SMBV-K23 (Pulse Generator)
[:SOURce<hw>]:PULM:WIDTh
<Width>
The command sets the width of the generated pulse. The width determines the pulse
length. The pulse width must be at least 20ns less than the set pulse period.
Parameters:
<Width>
float
Range:
10 ns to 100 s
Increment: 10 ns
*RST:
2 us
Example:
PULM:WIDT 33 us
sets a width of 33 us for the pulse.
Options:
R&S SMBV-K23 (Pulse Generator)
7.14.19 SOURce:ROSCillator Subsystem
This subsystem contains the commands for setting the external and internal reference
frequency.
The commands of this subsystem are not affected by an instrument reset (
..................................................................690
.............................................................691
.................................................................691
..............................................................691
............................................................692
.....................................................................................692
[:SOURce]:ROSCillator:EXTernal:FREQuency
<Frequency>
Selects the external reference frequency.
5 MHz can be only operated, if the instrument has an RF board starting from version
1406.7220.
Parameters:
<Frequency>
5MHZ | 10MHZ
*RST:
10MHZ
Example:
ROSC:SOUR EXT
selects the external source. The reference must be input at the
REF IN input.
ROSC:EXT:FREQ 10MHz
slects 10 MHz external reference frequency.
SOURce Subsystem
Содержание SMBV100A
Страница 31: ......
Страница 117: ...Manual Operation R S SMBV100A 100 Operating Manual 1407 6062 32 08 Legend of Front Panel Controls ...