Remote Control Commands
R&S
®
SMBV100A
565
Operating Manual 1407.6062.32 ─ 08
<RCosine>
float
Range:
0.05 to 1.0
Increment: 0.01
*RST:
0.35
<SPhase>
float
Range:
0.15 to 2.5
Increment: 0.01
*RST:
2
Example:
BB:DM:FILT:PAR:APCO25 0.2
sets the roll-off factor to 0.2 for filter type APCO25.
[:SOURce<hw>]:BB:DM:FILTer:PARameter:COSine:BANDwidth
<FiltParm>
Sets the bandwidth of the cosine filter.
Parameters:
<FiltParm>
float
Range:
400 to 25E6
Increment: 1E-3
*RST:
270833.333
Example:
:BB:DM:FILT:TYPE COS
:BB:DM:FILT:PAR:COS:BAND 7200
[:SOURce<hw>]:BB:DM:FILTer:PARameter:APCO25Lsm:GAUSs
<FiltParm>
[:SOURce<hw>]:BB:DM:FILTer:PARameter:APCO25Lsm:LOWPass
<FiltParm>
Sets the cut off frequency for the lowpass/ gauss filter.
Parameters:
<FiltParm>
float
Range:
400 to 25E6
Increment: 1E-3
*RST:
270833.333
Example:
:BB:DM:FILT:TYPE APCO25Lsm
:BB:DM:FILT:PAR:APCO25L:LOWP 5166.667
:BB:DM:FILT:PAR:APCO25L:GAUS 1700
[:SOURce<hw>]:BB:DM:FILTer:TYPE
<Type>
The command selects the filter type.
When a standard is selected (
:BB:DM:STAN
), the filter type and filter parameter are set