Instrument Function
R&S
®
SMBV100A
187
Operating Manual 1407.6062.32 ─ 08
From
Sets the start value of the index range.
SCPI command:
n.a.
Range
Sets the range for filling the table.
SCPI command:
n.a.
Select column to fill
Selects either the frequency or the level column to be filled with the value defined below.
SCPI command:
n.a.
Start value
Sets the start value for the frequency or the level entries.
SCPI command:
n.a.
End value
Displays the end value for the frequency or the level entries.
SCPI command:
n.a.
Increment value
Sets the increment for the frequency or the level entries.
SCPI command:
n.a.
Fill
Fills the selected column in the set range with values, starting with the start value and
using the set increment.
SCPI command:
n.a.
5.4.10.3
Filling the Correction List with Power Sensor Measurement Data
The level correction values for the user correction list can be acquired by means of R&S
NRP power sensors. The R&S NRP sensors are connected to either the SENSOR con-
nector or to one of the USB interfaces. Configuration of the connection is performed in
the "Power Sensor" menu (see
chapter 5.4.6, "NRP-Z Power Viewer"
, on page 163). The
filling of the ucor list with measurement data is performed in the ucor list editor (see
User Cor. Data - User Correction"
In the editor, the frequencies for which the correction values are to be acquired are
entered in the frequency column (either manually or by means of the "Fill..." menu).
RF Signal and Analog Modulations - RF Block
Содержание SMBV100A
Страница 31: ......
Страница 117: ...Manual Operation R S SMBV100A 100 Operating Manual 1407 6062 32 08 Legend of Front Panel Controls ...