![Rohde & Schwarz SMBV100A Скачать руководство пользователя страница 517](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_manual_1477734517.webp)
Remote Control Commands
R&S
®
SMBV100A
500
Operating Manual 1407.6062.32 ─ 08
Parameters:
<Mode>
ONLY | ADD | CW
ADD
The AWGN noise signal is added to the baseband signal.
ONLY
The pure AWGN noise signal is modulated to the carrier. The con-
nection to the baseband is interrupted
CW
The sine interfering signal is added to the baseband signal.
*RST:
ADD
Example:
AWGN:MODE ONLY
activates the generation of a pure noise.
AWGN:STAT ON
switches on the generation of a pure noise.
[:SOURce<hw>]:AWGN:POWer:CARRier
<Carrier>
available for modes "Additive Noise" and "CW Interferer" (
SOUR:AWGN:MODE ADD|CW
)
This command either sets or queries the carrier or signal level depending on the selected
reference mode.
●
"Reference Mode Carrier" (
SOUR:AWGN:POW:RMOD CARR
)
Sets the carrier level. The level of the noise signal is derived from the entered C/N
value.
●
"Reference Mode Noise" (
SOUR:AWGN:POW:RMOD NOIS
)
queries the carrier level which is derived from the entered C/N value. The noise level
is set with command
SOUR:AWGN:POW:NOISe
.
Parameters:
<Carrier>
float
Range:
min to 20
Increment: 0.01
*RST:
0
Default unit: dBm
Example:
AWGN:DISP:MODE RF
sets the display mode to RF
AWGN:POW:CARR?
queries the carrier level
[:SOURce<hw>]:AWGN:POWer:MODE
<Mode>
This command selects the mode for setting the noise level.
This command is available for mode "Additive Noise" (
SOUR:AWGN:MODE ADD
).
SOURce Subsystem
Содержание SMBV100A
Страница 31: ......
Страница 117: ...Manual Operation R S SMBV100A 100 Operating Manual 1407 6062 32 08 Legend of Front Panel Controls ...