Remote Control Commands
R&S
®
SMBV100A
673
Operating Manual 1407.6062.32 ─ 08
[:SOURce<hw>]:PM:SENSitivity?
The command queries the input sensitivity of the EXT MOD input in RAD/V. The com-
mand is only effective if the external modulation source is selected (
SOUR:PM:SOUR
EXT
). The returned value depends on the modulation deviation setting
(
SOUR:PM:DEViation
). This value is assigned to the voltage value for full modulation
of the input.
Return values:
<Sensitivity>
float
Example:
PM:DEV 1
sets a modulation deviation of 1RAD.
PM:SENS?
queries the input sensitivity at the EXT MOD input.
Response: 1
since the voltage value for full modulation is 1V, the resulting sen-
sitivity is precisely 1RAD/V.
Usage:
Query only
SCPI confirmed
[:SOURce<hw>]:PM:SOURce
<Source>
The command selects the modulation source for phase modulation.
Parameters:
<Source>
INTernal | EXTernal | INT,EXT
INTernal
INT is the LF generator. The frequency of the internal modulation
signal can be set in the
SOURce:LFOutput
subsystem.
EXTernal
The external signal is input at the EXT MOD connector.
INT,EXT
Internal and external modulation source can be selected at the
same time.
*RST:
INTernal
Example:
PM:SOUR INT
selects the internal modulation source.
Usage:
SCPI confirmed
[:SOURce<hw>]:PM:STATe
<State>
The command activates/deactivates phase modulation. Activation of phase modulation
deactivates frequency modulation.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
OFF
SOURce Subsystem