Remote Control Commands
R&S
®
SMBV100A
463
Operating Manual 1407.6062.32 ─ 08
:DISPlay:DIALog:OPEN
<DialogId>
Opens the specified dialog. To determine the dialog identifier, use command
.
Setting parameters:
<DialogId>
string
Example:
DISP:DIAL:OPEN "<dialog ID>"
opens the dialog, determined with the "<dialog ID>".
Usage:
Setting only
:DISPlay:PSAVe:HOLDoff
<Holdoff>
Sets the waiting time for the screen-save mode of the display. The entry is dimensionless.
Parameters:
<Holdoff>
integer
Range:
1 to 60
*RST:
10
Default unit: minute
Example:
DISP:PSAV:HOLD 8
sets the timeout of the screen saver to 8 minutes.
:DISPlay:PSAVe[:STATe]
<State>
Activates the screen-save mode of the display. If activated, the display including backlight
is completely switched off after the elapse of the wait time, provided no entries via front
panel, external mouse or external keyboard are made. To determine the wait time, use
command
.
This mode is recommended for protecting the display, especially if the instrument is
operated via remote control.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
OFF
Example:
DISP:PSAV ON
activates screen saver mode.
7.8 FORMat Subsystem
The FORMat subsystem contains the commands which determine the format of the data
that the R&S SMBV returns to the controller. This affects all query commands which
return a list of numerical data or block data. Reference is made to this in the descriptions
of the commands.
FORMat Subsystem
Содержание SMBV100A
Страница 31: ......
Страница 117: ...Manual Operation R S SMBV100A 100 Operating Manual 1407 6062 32 08 Legend of Front Panel Controls ...