Remote Control Commands
R&S
®
SMBV100A
669
Operating Manual 1407.6062.32 ─ 08
Parameters:
<Source>
AUTO | IMMediate | SINGle | BUS | EXTernal
AUTO|IMMediate
The trigger is free-running, i.e. the trigger condition is fulfilled con-
tinuously. The selected list in List mode is restarted as soon as it
is finished.
SINGle|BUS
The list is triggered by the GPIB commands
EXTernal
The list is triggered externally via the INST TRIG connector. The
list is executed once.
*RST:
SINGle
Example:
LIST:TRIG:SOUR EXT
selects triggering by means of the external trigger.
7.14.13 SOURce:MODulation Subsystem
This subsystem contains the command for switching on/off all modulations.
[:SOURce<hw>]:MODulation[:ALL]:STATe
<State>
The command switches the modulations on and off.
The command
SOUR:MOD:ALL:STAT OFF
switches all modulations off. A subsequent
command
SOUR:MOD:ALL:STAT ON
restores the status that was active before the last
switch-off. "MOD OFF" is displayed in the info line of the header next to the "Level" field.
Parameters:
<State>
0 | 1 | OFF | ON
Example:
MOD:STAT OFF
switches off all modulations.
7.14.14 SOURce:PGEN Subsystem
This subsystem contains the commands for setting the pulse generator.
........................................................................................669
[:SOURce<hw>]:PGEN:STATe
<State>
The command activates/deactivates the output of the video/sync signal at the PULSE
VIDEO connector at the rear of the instrument.
The signal output and the pulse generator are automatically switched on with activation
of pulse modulation if pulse generator is selected as modulation source. The signal output
can be switched off subsequently.
SOURce Subsystem