"Triangle"
One sweep runs from start to stop frequency and back, i.e. the shape
of the sweep resembles a triangle. Each subsequent sweep starts at
the start frequency.
SCPI command:
Step Lin/Log - Frequency Sweep
Sets the step width for the individual sweep steps. This entry is effective for all sweep
modes.
"Step Lin" or "Step Log" is displayed depending on whether "Spacing Lin" or "Log" is
selected.
"Step Lin"
With the linear sweep, the step width is a fixed frequency value which
is added to the current frequency. The linear step width is entered
in
Hz.
SCPI command:
"Step Log"
With the logarithmic sweep, the step width is a constant fraction of the
current frequency. The logarithmic step width is entered in
%.
Successive frequencies are calculated as follows:
●
start_f < stop_f
f2 = f1 * (1 + step_log / 100)
If f2 > stop_f: f2 is set to stop_f.
●
start_f > stop_f
f2 = f1 / (1 + step_log / 100)
If f2 < stop_f: f2 is set to stop_f.
When the shape "Triangle" is set, the frequency values on the slope
from
stop_f back to start_f
are the same as on the slope from
start_f
to stop_f
.
SCPI command:
Dwell Time - Frequency Sweep
Sets the dwell time. The dwell time determines the duration of the individual sweep steps.