Remote Control Commands
R&S
®
SMBV100A
655
Operating Manual 1407.6062.32 ─ 08
Parameters:
<Logarithmic>
float
Range:
0.01PCT to 100PCT
Increment: 0.01PCT
*RST:
1
Example:
LFO:FREQ:STAR
sets the start frequency to 1 kHz.
LFO:FREQ:STOP
sets the stop frequency to 100 kHz.
LFO:SWE:SPAC LOG
sets logarithmic sweep spacing.
LFO:SWE:STEP:LOG 10PCT
sets the step width for logarithmic sweep spacing to 10% of the
previous frequency in each instance.
[:SOURce]:LFOutput<ch>:SHAPe
<Shape>
The command selects the shape of the LF generator.
The instrument is equipped with one LF geneator only and the suffix has no impact.
Parameters:
<Shape>
SINE | SQUare
*RST:
SINE
Example:
LFO:SHAP SQU
selects a rectangular shape for the signal of the LF generator.
Usage:
SCPI confirmed
[:SOURce]:LFOutput:SIMPedance
<Simpedance>
The command selects the output impedance of the LF generator. Selection LOW and
600 Ohm are available.
For instruments with older version of the RF board, the LF output impedance is forced to
LOW and cannot be changed.
Parameters:
<Simpedance>
LOW | G600
*RST:
LOW|
Example:
SOUR:LFO:SIMP G600
'sets the output impedance of the LF generator to 600 Ohms
[:SOURce]:LFOutput:VOLTage
<Voltage>
The command sets the output voltage of the LF output.
SOURce Subsystem