Remote Control Commands
R&S
®
SMBV100A
630
Operating Manual 1407.6062.32 ─ 08
Usage:
SCPI confirmed
[:SOURce<hw>]:FREQuency:MODE
<Mode>
The command sets the instrument operating mode and therefore also the commands
used to set the output frequency.
Parameters:
<Mode>
CW | FIXed | SWEep | LIST
CW|FIXed
The instrument operates in fixed-frequency mode;
CW
and
FIXed
are synonyms. The output frequency is set
with
:SOURce:FREQuency:CW|FIXed
.
SWEep
The instrument operates in SWEep mode. The frequency is set
using the commands
SOURce:FREQuency:STARt; STOP;
CENTer; SPAN; MANual
.
LIST
The instrument processes a list of frequency and level settings for
the selected path.
The List mode settings are made in the
SOURce:FREQuency:MODE LIST
also sets the command
SOURce:POWer:MODE
automatically to
LIST
.
Correlation
:
FREQ:MODE LIST
sets
POW:MODE LIST
*RST:
CW
Example:
FREQ:MODE SWE
sets the SWEep mode. The settings under
SOURce:FREQuency:STARt
;
STOP
;
CENTer
;
SPAN
;
MANual
become effective.
Usage:
SCPI confirmed
[:SOURce<hw>]:FREQuency:MULTiplier
<Multiplier>
The command sets the value for the multiplication factor of a subsequent downstream
instrument.
Parameters:
<Multiplier>
float
Range:
1 to 10000
*RST:
1
Example:
FREQ:MULT 1
sets the multiplication factor to 1.
SOURce Subsystem