Remote Control Commands
R&S
®
SMBV100A
675
Operating Manual 1407.6062.32 ─ 08
Parameters:
<OffMode>
SHOLd | TABLe | HIGHaccuracy
SHOLd
Level control is activated briefly if the level or frequency changes
("ALC Off Sample & Hold").
HIGHaccuracy
The ALC mode High Accuracy ensures improved level accuracy.
Note:
This mode increases the frequency and level setting time
up to 20 ms!
TABLe
Internal level control is performed according to the ALC table.
*RST:
SHOLd
Example:
POW:ALC OFF
deactivates automatic level control for RF output A.
POW:ALC:OMOD SHOL
level control is briefly activated if the frequency or level changes.
[:SOURce<hw>]:POWer:ALC:SONCe
The command briefly activates level control for correction purposes.
Example:
POW:ALC OFF
deactivates automatic level control for RF output A.
POW:ALC:SONC
level control is performed once only.
Usage:
Event
[:SOURce<hw>]:POWer:ALC[:STATe]
<State>
The command activates/deactivates automatic level control.
Parameters:
<State>
0 | OFF | AUTO | 1 | ON | PRESet | OFFTable | ONSample |
ONTable | OFFHighAccuracy
ON
Internal level control is permanently activated.
OFF|ONSample
Internal level control is deactivated.
AUTO
Internal level control is activated/deactivated automatically
depending on the operating state.
ONTable
Internal level control is performed according to the ALC table.
OFFHighAccuracy
The ALC mode High Accuracy ensures improved level accuracy.
Note:
This mode increases the frequency and level setting time
up to 20 ms!
SOURce Subsystem