Remote Control Commands
R&S
®
SMBV100A
639
Operating Manual 1407.6062.32 ─ 08
Parameters:
<Angle>
float
Range:
-10.0DEG to 10.0DEG
Increment: 0.02DEG
*RST:
0DEG
Default unit: DEG
Example:
IQ:IMP:QUAD:ANGL -5DEG
sets the quadrature offset to -5 degrees.
[:SOURce<hw>]:IQ:IMPairment[:STATe]
<State>
The command activates (ON) and deactivates (OFF) the three impairment or correction
values LEAKage, QUADrature and IQRatio for the analog signal in the I/Q modulator.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
OFF
Example:
IQ:IMP OFF
deactivates I/Q impairment.
[:SOURce<hw>]:IQ:SOURce
<Source>
This command selects the input signal for the I/Q modulator.
Selecting
ANALog
(Wideband I/Q In) switches off:
●
SOURce:AM:STATe OFF
●
SOURce:BBAM:STATe OFF
●
SOURce:BB:DM:STATe OFF
●
SOURce:BB:<Dig. Standard>:STATe OFF
Parameters:
<Source>
BASeband | ANALog
*RST:
BASeband
Example:
IQ:SOUR ANAL
selects an external analog signal as the input signal. The signal
must be applied at the inputs I and Q.
[:SOURce<hw>]:IQ:STATe
<State>
This command activates and deactivates I/Q modulation.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
OFF
Example:
IQ:STAT ON
activates I/Q modulation.
SOURce Subsystem
Содержание SMBV100A
Страница 31: ......
Страница 117: ...Manual Operation R S SMBV100A 100 Operating Manual 1407 6062 32 08 Legend of Front Panel Controls ...