Instrument Function
R&S
®
SMBV100A
308
Operating Manual 1407.6062.32 ─ 08
Hex ← Data List Editor
Switchover to hexadecimal display.
Each four bits are displayed as a hexadecimal value: To increase readability, the hexa-
decimal values in turn are displayed in pairs of two. The hex functions are automatically
assigned to the numeric keys at the front panel.
SCPI command:
n.a.
Save (Data List) ← Data List Editor
Saves the changes made to the Data List file selected for editing.
SCPI command:
n.a.
Control and Marker List Editor
The "Control and Marker List Editor" for editing the selected control list is called up in the
"List Management" submenu of the "Custom Digital Modulation" main menu by means
of the "Edit Control List"... button.
The available marker signals and the CW, Hop, Burst Gate and Level Attenuation control
signals can be defined in the "Control and Marker List Editor".
The marker signals defined in the Control and Marker List Editor are activated by selecting
the Marker mode "CList" in the "Trigger/Marker/Clock" menu. The control signals have
to be activated by selecting an "Internal Source" in the respective setting menu, e.g. the
"Power Ramp Control" menu for the Burst Gate and Level Attenuation control signals.
The available marker/control signals are color-coded. In the left "Configure Control Sig-
nal" section, each individual signal is assigned a colour; a check in the check box shows
the marker for which the "CList" marker type has been selected and the control signal for
which the "Internal" source has been selected.
Baseband Signal - Baseband Block