Remote Control Commands
R&S
®
SMBV100A
649
Operating Manual 1407.6062.32 ─ 08
In Sweep mode (
SOUR:LFO:FREQ:MODE SWE
), the frequency is linked to the sweep
frequency.
Parameters:
<Frequency>
float
Range:
0.1 Hz to 1 MHz
Increment: 0.1 Hz
*RST:
1 kHz
Default unit: Hz
Example:
LFO:FREQ 5kHz
sets the frequency of the LF signal to 5 kHz
[:SOURce]:LFOutput:FREQuency:MANual
<Manual>
In Sweep mode (
SOUR:LFO:FREQ:MODE SWE
) the command sets the frequency for the
next sweep step in the "Step" sweep mode (
SOUR:LFO:SWE:MODE MAN
). Here only
frequency values between the settings
SOUR:LFO:FREQ:STAR
and
...:STOP
are per-
mitted. Each sweep step is triggered by a separate
SOUR:LFO:FREQ:MAN
command.
Parameters:
<Manual>
float
Range:
STARt to STOP
Increment: 0.1 Hz
*RST:
1 kHz
Example:
LFO:SWE:MODE MAN
sets the "Step" sweep mode.
LFO:FREQ:MAN 5 kHz
sets an LF frequency of 5 kHz for the next step in the "Step" sweep
mode.
LFO:FREQ:MODE SWE
sets the LF Sweep mode. An LF frequency of 5 kHz is output.
LFO:FREQ:MAN 5.1 kHz
triggers the next sweep step with a frequency of 5.1 kHz.
[:SOURce]:LFOutput:FREQuency:MODE
<Mode>
The command sets the instrument operating mode and therefore also the commands
used to set the output frequency.
SOURce Subsystem
Содержание SMBV100A
Страница 31: ......
Страница 117: ...Manual Operation R S SMBV100A 100 Operating Manual 1407 6062 32 08 Legend of Front Panel Controls ...