Instrument Function
R&S
®
SMBV100A
107
Operating Manual 1407.6062.32 ─ 08
Adjustment for only the set frequency is considerably faster than adjustment across the
entire frequency range. This is possible with "Full Range".
SCPI command:
Adjust I/Q Modulator Full Range
Starts the adjustment for the I/Q modulator for the entire frequency range. The I/Q mod-
ulator is adjusted with respect to carrier leakage, I/Q imbalance and quadrature.
SCPI command:
5.2.3.2
Hardware Config
In the "Hardware Config" dialog, the installed assemblies together with their variants and
revision states can be displayed for servicing purposes.
To open the "Hardware Config" dialog, select "System" and press the SETUP or
MENU key.
Section "Counter" in the upper part of the menu shows the "Operation Time" (in hours)
and the number of power-on ("Power On Counter").
The second part of the menu is a table that lists the installed assemblies. It is divided into
the sections:
●
"Common Assembly"
●
"RF Assembly"
●
"Baseband Assembly"
Operation Time / h
Displays the operation time in hours.
SCPI command:
Power On Count
Displays the number of power-on.
SCPI command:
General Instrument Settings
Содержание SMBV100A
Страница 31: ......
Страница 117: ...Manual Operation R S SMBV100A 100 Operating Manual 1407 6062 32 08 Legend of Front Panel Controls ...