![Rohde & Schwarz SMBV100A Скачать руководство пользователя страница 388](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_manual_1477734388.webp)
Instrument Function
R&S
®
SMBV100A
371
Operating Manual 1407.6062.32 ─ 08
The buttons in the lower part of the menu open dialogs for defining carrier tables and for
configuring markers, triggers and clocks.
The carrier settings can be visualized in the graphical "Carrier Graph" dialog.
General Settings for Multi Carrier CW
The upper part of the menu is used for powering up the Multi Carrier CW signal and calling
the default settings.
State
Enables/disables the Multi Carrier CW.
Switching on this standard turns off all the other digital standards and digital modulation
modes.
Since Multi Carrier CW signals are computed in arbitrary waveform mode, changes to
the settings of individual carriers are not adopted until the Accept button is pressed. This
applies to the settings in the "Table Setup Assistant" section and the "Carrier Table"
dialog.
SCPI command:
Set to Default
Calls default settings. The values are shown in the following table.
Parameter
Value
State
Not affected by "Set to Default"
Carrier Setup
Number of Carriers
64
Carrier Spacing
10 kHz
Optimize Crest Factor
Chirp
Baseband Signal - Baseband Block