Remote Control Commands
R&S
®
SMBV100A
725
Operating Manual 1407.6062.32 ─ 08
:UNIT:ANGLe
<Angle>
The command defines the default unit for the phase modulation angle. It is not valid for
other commands which determine angle values, e.g. RF phase. It does not influence the
manual control parameter unit and the display.
Parameters:
<Angle>
DEGree | DEGRee | RADian
*RST:
RADian
Example:
UNIT:ANGL DEG
sets default unit DEG for all commands which determine angle
values.
:UNIT:POWer
<Power>
The command defines the default unit for power. It is valid for all commands which deter-
mine power values. It does not influence the manual control parameter unit and the dis-
play.
Parameters:
<Power>
V | DBUV | DBM
*RST:
DBM
Example:
UNIT:POW V
sets default unit V for all commands which determine power val-
ues.
UNIT Subsystem