10.7.3.2
STEP Transients
STEP transients allow a ramp or stairstep output profile to be programmed without creating LIST
transients with multiple entries. In STEP transient mode, the user enters initial value, end value,
delta step size, step duration, step count and no of repeats.
The STEP EXAMPLE shows the relationship between the various parameters.
This information is transformed to a standard Segment based transient list by the power source
when the LOAD button is pressed.
During transient execution, a progress indicator is visible showing percent completion.