LMX SERIES™ OPERATION MANUAL
SECTION 8:
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
LMX Series Power Source Operation Manual
Page 236 of 463
8.6.2
Execution Commands
The following commands may be used to control stored program executions.
Command Syntax
PROGram[:SELected]:EXECute [<nr1> [,<cr>]
Description
This command executes the selected program (default) or the
program number passed as the first parameter.
Parameters
Option: PROGRAM NUMBER [,<nr1>]
Option: MEMORY [,<cr>]– Available are: INTERNAL | RAM | USB
Parameter Format
[<nr1> [,<cr>]
Example
PROG:EXEC 2
Query Format
PROGram[:SELected]:EXECute ?
Description
The query format returns the number of the program that is in effect.
Parameter Format
<nr1>
Example
PROG:EXEC?
23
Command Syntax
PROGram:EXECute:TRANS [<nr1> [,<cr>]
Description
This command executes the selected program’s (default) transient
table or the program number passed as the first parameter.
Parameters
Option: PROGRAM NUMBER [,<nr1>]
Option: MEMORY [,<cr>]– Available are: INTERNAL | RAM | USB
Parameter Format
[<nr1> [,<cr>]
Example
PROG:EXEC:TRANS 2
Command Syntax
PROGram:EXECute:TRANS:RESET
Description
This command serves the same purpose as the “PROGram:TRANsient
STOP” command. See next.
This command resets the active transient execution. This command is
included to support UPC compatibility mode although it is not
documented in the UPC manual. Not recommended for new
programs.
Parameters
None
Parameter Format
n/a
Example
PROG:EXEC:TRANS:RESET