5-45
5.6.12 Backup Commands
RAC
Read the entire backup memory
Reads only the currently selected CONDITION number.
SAC
Writes data to the entire backup memory
Writes data only to the currently selected CONDITION number.
ACCn
Backup memory initialization control
n = 0: Initializes all items.
n = 1: Initializes all items excluding the analog calibration data specified by CAL. (GP-IB
settings are not initialized.)
5.6.13 Bus Commands (Applied Only to GP-IB)
GTL (Go To Local)
Cancels the remote mode and accepts key operations from the panel key.
Effective also in the local lockout.
SDC (Selected Device Clear)
Performs the same function as IFC. The GP-IB interface is initialized to the condition when
the power is turned ON. Since this command is an addressed command, only the TS-3200A
can be cleared out of multiple devices connected to the GP-IB line.
GET (Group Execute Trigger)
This command is applied when the GATE setting is EXT.
The TRIG IN signal of the EXT GATE function, for which TRIG IN setting is required.