5.6.8 Analog Output Settings
DASc(,d)
D/A output source
c = 1 to 2: 3 or greater is ignored.
d = 0:TORQUE
1:REVO
2:POWER
DADc(,d)
D/A output data
c = 1 to 2: 3 or greater is ignored.
d = 0.1 to 10.00 [V]
Output full scale setting
DABc(,d)
D/A output method
If the TS-3200A receives only VOLTAGE, it ignores this command.
c = 1 to the number of channels mounted
d = 0:VOLTAGE(0-10V)
1 or greater is ignored.
DATc(,d)
D/A output time constant
c = 1 to 2: 3 or greater is ignored.
d = 0:OFF, 1:16ms, 2:31ms, 3:63ms, 4:125ms, 5:250ms, 6:500ms, 7:1s, 8:2s, 9:4s,
A:8s,
B:16s,C:32s,D:64s
0:OFF can be selected only when the TS-0321 is mounted and 4ms SAMPLE is
selected.
ABSn(d)
Absolute mode
n = 4:ANALOG OUT CH1
5:ANALOG OUT CH2
d = 0:NOR (usual), 1:ABS (absolute value)