5.1
Analog Output High-Speed Response (1ms) TS-0321A
The TS-3200A has a standard time constant of 16ms to 64s for the analog output.
When the TS-0321A option is built in the TS-3200A, 4ms or 1ms can be selected as the sampling time.
When 1ms sampling time is selected, the analog output is updated at 1ms sampling intervals allowing
high-speed response.
When the TS-0321A is built in and 4ms sampling time is selected, OFF (exponential averaging not
performed) is added to the time constant setting for the analog output.
However, there are the following restrictions for 1ms sampling time.
Restrictions
●
The comparator function (CMP 1 OUT, CMP 2 OUT, CMP 3 OUT and CMP 4 OUT
terminals) does not operate.
●
The time constant for the analog output cannot be set.
Only the ON/OFF setting of the low-pass filter (fc/80Hz40 dB/dec) can be made.